Effect on CT value of peripheral tissue with different concentrations of #br#
contrast media using simulated vascular enhancement |
LI Wei-wei |
Tai’an Central Hospital Branch, Tai’an Shandong 271000, China |
|
|
Abstract Objective: To study the influence of contrast agent on CT value of tissue around syringe with different concentrations of contrast agent in the paraffin mold. Methods: Paraffin phantom of syringe with 11 different concentrations of contrast agent and saline as control was scanned with the same condition, and the CT value of the study area was measured in pixels and analyzed. Results: Contrast concentration had impact on CT value of different areas. CT beam hardening effect decreased the CT value of the small area around blood vessels with the increase of contrast agent concentration. Volume effect increased the CT value of area outside the beam hardening effect around blood vessels with the increase of contrast agent concentration. Conclusion: CT values of surrounding tissue increase because of volume effect. There is significant difference between each experimental group and control group(P<0.004). Decreases of CT value of peripheral tissue due to pixel tube beam hardening effect is not significant between each experimental group and control group(P>0.004).
|
Received: 03 September 2017
|
|
|
|
|
[1]张越,万业达,何岸苇,等. 不同管电压下周围结构对CT值测量影响的研究[J]. 天津医科大学学报,2013,19(3):199.
[2]彭文献,彭天舟,叶小琴,等. CT扫描参数对人体组织CT值影响的研究[J]. 中华放射医学与防护杂志,2010,30(1):79.
[3]兰永树,黄文龙,朱德强,等. 影响CT值测量的因素及对策的探讨[J]. 实用放射学杂志,2006,22(3):35.
[4]白人驹,徐克. 医学影像学[M]. 7版. 北京:人民卫生出版社,2015:12.
[5]赵传军,王烈伟. CT图像伪影及其解决方案[J]. 中国医疗设备,2014,29(4):166-167.
[6]杨克柽,王余锋. CT图像伪影[J]. 医疗装备,2006,19(5):16-17.
[7]柴佳彬. CT成像硬化及金属伪影校正算法研究[D]. 兰州:兰州大学,2016.
[8]陆皓,后军民,任静,等. CT图像中射线束硬化伪影的分析[J]. 中国中西医结合影像学杂志,2015,13(6):711-712.
|
|
|
|