双层探测器能谱CT单能级成像对门静脉成像质量的影响
王世伟1,李金锋2,邹  颖1,李  涛2,罗春材2,黄自立2,李雪萍2
1. 解放军总医院海南医院放射科,海南 三亚  572013;2. 解放军总医院第一医学中心放射科,北京  100853
Improving image quality in portal venography with dual-layer detector spectral CT imaging
WANG Shi-wei1, LI Jin-feng2, ZOU Ying1, LI Tao2, LUO Chun-cai2, HUANG Zi-li2, LI Xue-ping2
1. Department of Radiology, PLA General Hospital, Hainan Hospital, Sanya Hainan 572013, China;
2. Department of Radiology, First Medical Center, PLA General Hospital, Beijing 100853, China
全文: PDF (0 KB)   HTML (1 KB) 
输出: BibTeX | EndNote (RIS)      
摘要 目的:探索双层探测器能谱CT单能级影像对于门静脉成像质量的影响。方法:使用单源双层探测器能谱CT对39例怀疑腹部疾病的患者行对比增强检查。以3 mL/s的速度注射对比剂iopamidol(320 mg/mL)60 mL后60 s以能谱CT采集方式行门脉期扫描。原始数据被重建成常规120 kVp的混合能量图像及以能谱为基础的数据,获得40,50,60,70 keV的单能级图像。测量常规120 kVp混合能量组以及40,50,60,70 keV的单能级组的图像噪声,门静脉的信噪比(SNR),门静脉与同层竖脊肌相比较的对比噪声比(CNR)。统计方法采用方差分析比较各组之间影像噪声、SNR和CNR,寻找门静脉成像的最佳单能级图像。并对图像质量进行主观评分。结果:各组之间影像噪声、门静脉的SNR、与竖脊肌相比较的CNR有显著的统计学差异(P<0.05)。40 keV组的噪声低于常规120 kVp混合能量组,SNR和CNR远远高于常规120 kVp混合能量组及其他单能量组。主观评分显示40 keV最佳单能级图像质量优于常规混合能量图像质量,两个医师评分的一致性较好。结论:双层探测器能谱CT 40 keV单能级成像显示门脉成像质量最佳。
服务
把本文推荐给朋友
加入我的书架
加入引用管理器
E-mail Alert
RSS
作者相关文章
王世伟1
李金锋2
邹 颖1
李 涛2
罗春材2
黄自立2
李雪萍2
关键词 门静脉;体层摄影术X线计算机    
Abstract:Objective: To investigate the effect of optimal monochromatic energy image on the image quality of portal venography with dual-layer detector spectral CT. Methods: Thirty-nine patients suspected of abdominal disease underwent contrast-enhanced CT examination using a single-source, dual-detector spectral CT. The portal vein phase scan was performed 60 seconds after 60 mL iopamidol(320 mg/mL) injection at the speed of 3 mL/s. The raw images were reconstructed to obtain polychromatic 120 kVp images and spectral based images. The 40, 50, 60, 70 keV monochromatic images were gotten. Image noise, signal-to-noise ratio(SNR) of portal vein and contrast-to-noise ratio(CNR) of portal vein relative to the erector spine muscle from polychromatic 120 kVp images and 40, 50, 60, 70 keV monochromatic images were measured. Anova test was used to compare the noise, SNR and CNR among 5 groups. An optimal monochromatic image set was chosen for obtaining the best SNR and CNR for portal vein. Image quality was also evaluated on a five-point scale by two radiologists. Results: The difference of noise, SNR, CNR of portal vein among five groups were statistically significant(P<0.05). The noise at 40 keV group was significantly lower than that of polychromatic 120 kVp group. The SNR and CNR of 40 keV group was significantly higher than that of other groups(P<0.05), much higher than that of polychromatic 120 kVp group. The image quality score of 40 keV optimal monochromatic image was superior to that of polychromatic images. The agreement of image quality grading between two doctors was good. Conclusion: 40 keV is the optimal monochromatic image for displaying portal vein using dual-layer detector spectral CT.
Key wordsPortal vein    Tomography, X-ray computed
收稿日期: 2019-11-11     
PACS:  R331.34  
  R814.42  
通讯作者: 李涛,解放军总医院第一医学中心放射科,100853。E-mail:litaofeivip@163.com   
作者简介: 王世伟(1986-),男,海南屯昌人,技师。E-mail:404785375@qq.com
引用本文:   
王世伟1,李金锋2,邹 颖1,李 涛2,罗春材2,黄自立2,李雪萍2. 双层探测器能谱CT单能级成像对门静脉成像质量的影响[J]. 中国临床医学影像杂志, 2020, 31(6): 425-428.
WANG Shi-wei1, LI Jin-feng2, ZOU Ying1, LI Tao2, LUO Chun-cai2, HUANG Zi-li2, LI Xue-ping2. Improving image quality in portal venography with dual-layer detector spectral CT imaging. JOURNAL OF CHINA MEDICAL IMAGING, 2020, 31(6): 425-428.
链接本文:  
http://www.jccmi.com.cn/CN/10.12117/jccmi.2020.06.012     或     http://www.jccmi.com.cn/CN/Y2020/V31/I6/425
  友情链接
版权所有 © 2015 《中国临床医学影像杂志》编辑部
地址:沈阳市和平区三好街36号 邮编:110004 电话/传真: 024-23925069 电子信箱: jccmisy@sina.cn
本系统由北京玛格泰克科技发展有限公司设计开发