Abstract:Objective: To assess the diagnostic value of ultrasonographic elastography for identifying suspected malignant breast neoplasms. Materials and Methods: Eighty-five patients with small single solid breast neoplasms first underwent two-dimensional, color Doppler and power Doppler ultrasound. Then elastography was performed, and used an improved 5-point system to diagnose malignant or benign lesions. The ultrasonographic results were compared with pathologic findings. Results: Taking the elasticity score 4~5 as malignant standard, the diagnostic sensitivity was 89.18%, specificity was 89.58%, accuracy was 89.41%. The combined diagnostic sensitivity of two-dimensional, color Doppler and power Doppler ultrasound was 48.65%, specificity was 79.16%, accuracy was 65.88%. Conclusion: Real-time ultrasound elastography is a promising imaging technique that can assist in the differential diagnosis of malignant small single solid breast neoplasm.
梁 霞;吴泽惠;杨通明;唐石初;王志远;秦 昂. 实时超声弹性成像对乳腺单发实性小肿块的诊断价值[J]. , 2011, 22(4): 229-231.
LIANG Xia;WU Ze-hui;YANG Tong-ming;TANG Shi-chu;WANG Zhi-yuan;QIN Ang. The value of real-time ultrasound elastography in the differential diagnosis of small single solid breast neoplasms. , 2011, 22(4): 229-231.